INTENDED USE

ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) is a rapid chromatographic immunoassay for the qualitative detection of Antibody to Hepatitis B Surface Antigen in serum or plasma.



PERFORMANCE CHARACTERISTICS

Sensitivity

ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) has been tested with a sensitivity panel ranging from 1mIU/ml to 40mIU/ml. The test can detect 10mIU/ml of HBsAb in 15 minutes.

Specificity

Antigen used for ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) is highly specific for detecting HBsAb in serum or plasma. The specificity was comparable to RIA. 
 

Method

RIA

Total Results

ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma)

Results

Positive

Negative

Positive

231

3

234

Negative

0

180

180

Total Results

231

183

414

Relative Sensitivity: >99.9% (97.5%CI*: 98.7%-100% 

Relative Specificity: 98.4% (95%CI*: 95.3%-99.7%)  

Accuracy: 99.3% (95%CI*: 97.9%-99.9%)                    

*Confidence Intervals  

Precision  

Intra-Assay

Within-run precision has been determined by using 15 replicates of three specimens containing negative, low positive and high positive. The negative and positive values were correctly identified 99% of the time.

Inter-Assay

Between-run precision has been determined by using the same three specimens of negative, low positive and high positive of HBsAb in 15 independent assays. Three different lots of ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) has been tested over a 3-month period using negative, low positive and high positive specimens. The specimens were correctly identified 99% of the time.

Cross-reactivity

ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) has been tested by HAMA, Rheumatoid factor (RF), HAV, Syphilis, HIV, H. Pylori, MONO, CMV, Rubella and TOXO positive specimens. The results showed no cross-reactivity

Interfering Substances

ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) has been tested for possible interference from visibly hemolyzed and lipemic specimens. No interference was observed.

In addition, no interference was observed in specimens containing up to 2,000 mg/dL Hemoglobin, 1000 mg/dL Bilirubin, and 2000 mg/dL human serum Albumin.


 

LIMITATIONS

  1. ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) is for in vitro diagnostic use only. This test should be used for the detection of HBsAb in serum or plasma specimen.

  2. ACCU-TELL® Rapid HBsAb Test Strip (Serum/Plasma) cannot detect less than 10mIU/ml of HBsAb in specimens.

  3. As with all diagnostic tests, all results must be considered with other clinical information available to the physician.